A New Configuration for Helmet Continuous Positive Airway Pressure Allowing Tidal Volume Monitoring
2020
We performed a bench and human study to test the hypothesis that a ventilator can accurately estimate Vt when a helmet is used in CPAP mode in a single limb configuration with an intentional leak port placed at the helmet expiratory port. We have recently demonstrated the effectiveness of this setup in bilevel mode
Effectiveness and safety of a new helmet CPAP configuration allowing tidal volume monitoring in patients with COVID-19.
2021
Abstract Background High generated tidal volumes (Vt) have been correlated with higher risk of self-induced lung injury and worse clinical outcome. This study aimed to evaluate the effectiveness and safety of a new helmet continuous positive airway pressure delivered (h-CPAP) configuration allowing Vt monitoring in patients affected by COVID-19. Methods This prospective observational study was performed in the respiratory intermediate care unit of University Hospital in Turin, Italy, between March 24th, and June 15th, 2020. Included patients were treated with CPAP via a single-limb intentional leak configuration by a turbine-driven ventilator, provided with a dedicated patch. Effectiveness …